Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 863: (1 year) CAL VER RS-232C /5520,5139,85 DATE: 21-Aug-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 57 NUMBER OF LINES: 555 CONFIGURATION: Fluke 5520A STANDARD: Fluke PM 5139 STANDARD: Fluke 85 ============================================================================= # # Source: # Fluke 86X Performance Test Procedure, Preliminary, February 1995 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # ini_val.exe Executable: Gets default port for UUT from METCAL.INI # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # # The following describes the use of the [I] and [I$] special constructions # used in the PORT FSC in this procedure: # # QM (Query Measurement) command # # 3.001 PORT [D2000]QM[13][I][I][I][I$][I$] # # 1st [I] reads the acknowledge response # 2nd [I] reads the binary data count # 3rd [I] reads the primary display value # 1st [I$] reads the 1st secondary display value if present # 2st [I$] reads the 2nd secondary display value if present # # SF (Set Function) command # # 6.003 PORT SF 9[13][I][D1000] # # [I] reads the acknowledge response # # # Note, this procedure must be run with ASK- A, no termination to post test # on PASS or FAIL. This is because if ASK- F or ASK+ A is used and the # operator selects Repeat from post test, remote commands be re-sent via # the PORT FSC will toggle the UUT out of the correct mode for the test. # For example if the remote command "SF C" is sent to select AC Current, and # the Repeat option was allowed to be selected, the "SF C" command would be # re-sent and the UUT would be put in DC Current mode. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R Q N P A W 1.002 STD Fluke PM 5139 1.003 STD Fluke 85 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP No warm-up time requirement. 1.006 DISP Ambient temperature: 18C - 28C. 1.006 DISP Relative humidity: less than 90%. 1.007 DOS ini_val metcal.ini startup port 1.008 DISP Connect the UUT to [MEM2]. 1.009 DISP Rotate the UUT function switch to SET UP. 1.009 DISP 1.009 DISP Next Item (Soft Key 1) and left and right arrow keys 1.009 DISP (Soft Keys 2 & 3) to select: 1.009 DISP [32] Trigger Single 1.009 DISP [32] Baud Rate 4800 1.009 DISP [32] Parity None 1.010 PORT [P4800,N,8,1,X][T5000] 1.011 HEAD {RMS AC VOLTAGE TESTS} 1.012 DISP Connect the 5520A to the UUT as follows: 1.012 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode 1.012 DISP [32] 5520A NORMAL LO to UUT COM 1.013 DISP Rotate the UUT function switch to VAC. 1.013 DISP 1.013 DISP Select Advance when the numerical display appears. 1.014 HEAD RMS AC VOLTAGE TESTS: {300mV Range} 1.015 5520 30.00mV 20H SI S 2W 1.016 PORT [D2000]QM[13][I][I][I][I$] 1.017 MEME 1.018 MEMC 300 mV 0.55U 20H 2.001 5520 290.00mV 30kH SI S 2W 2.002 PORT [D2000]QM[13][I][I][I][I$] 2.003 MEME 2.004 MEMC 300 mV 1.55U 30kH 3.001 5520 100.00mV 100kH SI S 2W 3.002 PORT [D2000]QM[13][I][I][I][I$] 3.003 MEME 3.004 MEMC 300 mV 6.00U 100kH 4.001 5520 200.00mV 200kH SI S 2W 4.002 PORT [D2000]QM[13][I][I][I][I$] 4.003 MEME 4.004 MEMC 300 mV 18.00U 200kH 5.001 5520 50.00mV 300kH SI S 2W 5.002 PORT [D2000]QM[13][I][I][I][I$] 5.003 MEME 5.004 MEMC 300 mV 7.00U 300kH 6.001 HEAD RMS AC VOLTAGE TESTS: {3V Range} # Press the RANGE button. 6.002 PORT SF 0[13][I][D1000] # Press Range up (Soft Key 2) to select the 3V range. 6.003 PORT SF 9[13][I][D1000] 6.004 5520 0.1500V 20H SI S 2W 6.005 PORT [D2000]QM[13][I][I][I][I$] 6.006 MEME 6.007 MEMC 3 V 0.0033U 20H 7.001 5520 1.0000V 300kH SI S 2W 7.002 PORT [D2000]QM[13][I][I][I][I$] 7.003 MEME 7.004 MEMC 3 V 0.1200U 300kH 8.001 HEAD RMS AC VOLTAGE TESTS: {Peak Hold} 8.002 5520 2.828Vp 10kH SI S 2W # Press the RANGE button. 8.003 PORT SF 0[13][I][D1000] # Select Peak Hold (Soft Key 2). 8.004 PORT SF 9[13][I][D1000] 8.005 PORT [D2000]QM[13][I][I][I$][I][I$] 8.006 MEME 8.007 MEMC 3 Vp 0.342U 10kH # De-select Peak Hold. 9.001 PORT SF 9[13][I][D1000] 9.002 HEAD RMS AC VOLTAGE TESTS: {30V Range} # Press the RANGE button. 9.003 PORT SF 0[13][I][D1000] # Press up (Soft Key 2) to select the 30V range. 9.004 PORT SF 9[13][I][D1000] 9.005 5520 29.000V 10kH SI S 2W 9.006 PORT [D2000]QM[13][I][I][I][I$] 9.007 MEME 9.008 MEMC 30 V 0.155U 10kH 10.001 5520 3.200V 200kH SI S 2W 10.002 PORT [D2000]QM[13][I][I][I][I$] 10.003 MEME 10.004 MEMC 30 V 4% 0.200U 200kH 11.001 HEAD RMS AC VOLTAGE TESTS: {300V Range} # Press Range up (Soft Key 2) to select the 300V range. 11.002 PORT SF 9[13][I] 11.003 5520 15.00V 20H SI S 2W 11.004 PORT [D2000]QM[13][I][I][I][I$] 11.005 MEME 11.006 MEMC 300 V 0.33U 20H 12.001 5520 290.00V 50kH SI S 2W 12.002 PORT [D2000]QM[13][I][I][I][I$] 12.003 MEME 12.004 MEMC 300 V 10.7U 20kH 13.001 HEAD RMS AC VOLTAGE TESTS: {1000V Range} # Press Range up (Soft Key 2) to select the 1000V range. 13.002 PORT SF 9[13][I][D1000] 13.003 5520 1000.00V 50H SI S 2W 13.004 PORT [D2000]QM[13][I][I][I][I$] 13.005 MEME 13.006 MEMC 1000 V 16.0U 50H 14.001 5520 * S 14.002 HEAD {AVG AC VOLTAGE TESTS} # Press the RANGE button. 14.003 PORT SF 0[13][I][D1000] # Select Average (Soft Key 3). 14.003 PORT SF A[13][I][D1000] 14.004 HEAD AVG AC VOLTAGE TESTS: {300mV Range} 14.005 5520 0.00mV S 2W 14.006 PORT [D10000]QM[13][I][I][I][I$] 14.007 MEME 14.008 MEMC 300 mV 0.40U 15.001 5520 30.00mV 30kH SI S 2W 15.002 PORT [D3000]QM[13][I][I][I][I$] 15.003 MEME 15.004 MEMC 300 mV 0.60U 30kH 16.001 5520 290.00mV 50kH SI S 2W 16.002 PORT [D3000]QM[13][I][I][I][I$] 16.003 MEME 16.004 MEMC 300 mV 6.20U 50kH 17.001 HEAD AVG AC VOLTAGE TESTS: {1000V Range} # Press the RANGE button. 17.002 PORT SF 0[13][I][D1000] # Press Range up (Soft Key 2) four times to select the 1000V range. 17.003 PORT SF 9[13][I][D1000] 17.003 PORT SF 9[13][I][D1000] 17.003 PORT SF 9[13][I][D1000] 17.003 PORT SF 9[13][I][D1000] 17.004 5520 1000.00V 1kH SI S 2W 17.005 PORT [D3000]QM[13][I][I][I][I$] 17.006 MEME 17.007 MEMC 1000 V 19U 1kH 18.001 5520 * S 18.002 HEAD {DC VOLTAGE} 18.003 HEAD DC VOLTAGE TESTS: {30V Range} 18.004 DISP Rotate the UUT function switch to VDC. 18.004 DISP 18.004 DISP Select Advance when the numerical display appears. 18.005 5520 29.00V S 2W 18.006 PORT [D2000]QM[13][I][I][I][I$] 18.007 MEME 18.008 MEMC 30 V 0.009U 19.001 HEAD DC VOLTAGE TESTS: {300V Range} 19.002 5520 -290.00V S 2W 19.003 PORT [D2000]QM[13][I][I][I][I$] 19.004 MEME 19.005 MEMC 300 V 0.09U 20.001 HEAD DC VOLTAGE TESTS: {1000V Range} 20.002 5520 1000.00V S 2W 20.003 PORT [D2000]QM[13][I][I][I][I$] 20.004 MEME 20.005 MEMC 1000 V 0.50U 21.001 HEAD DC VOLTAGE TESTS: {Rel Mode} 21.002 5520 0.0V S 2W # Select Rel (Soft Key 1). 21.003 PORT SF 8[13][I][D1000] 21.004 5520 1000.00V S 2W 21.005 PORT [D2000]QM[13][I][I][I][I$] 21.006 MEME 21.007 MEMC V 0.40U 22.001 5520 * S # De-select Rel (Soft Key 1). 22.002 PORT SF 8[13][I][D1000] 22.003 HEAD DC VOLTAGE TESTS: {Touch Hold} 22.004 5520 0.0V S 2W # Press the TOUCH HOLD button. 22.005 PORT SF 1[13][I][D1000] 22.006 PORT [D2000]QM[13][I][I][I][I$] 22.007 MEME 22.008 MEMC V 0.2U 23.001 5520 * S 23.002 HEAD DC VOLTAGE TESTS: {300mV Range} 23.003 DISP Rotate the UUT function switch to mVDC. 23.003 DISP 23.003 DISP Select Advance when the numerical display appears. 23.004 5520 0.00mV S 2W 23.005 PORT [D2000]QM[13][I][I][I][I$] 23.006 MEME 23.007 MEMC 300 mV 0.02U 24.001 5520 15.00mV S 2W 24.002 PORT [D2000]QM[13][I][I][I][I$] 24.003 MEME 24.004 MEMC 300 mV 0.02U 25.001 HEAD DC VOLTAGE TESTS: {3000mV Range} 25.002 5520 -2900.00mV S 2W 25.003 PORT [D2000]QM[13][I][I][I][I$] 25.004 MEME 25.005 MEMC 3000 mV 0.9U 26.001 5520 * S 26.002 HEAD {DIODE TESTS} 26.003 DISP Rotate the UUT function switch to Diode/Capacitance. 26.003 DISP 26.003 DISP Select Advance when the numerical display appears. 26.004 HEAD DIODE TESTS: {Auto} 26.005 5520 2.90V S 2W 26.006 PORT [D2000]QM[13][I][I][I] 26.007 MEME 26.008 MEMC V 0.30U 27.001 5520 -2.90V S 2W 27.002 MATH MEM1 = -1 * MEM1 27.003 PORT [D2000]QM[13][I][I][I] 27.004 MEME 27.005 MEMC V 0.30U 28.001 HEAD DIODE TESTS: {Manual} # De-select Auto Diode (Soft Key 4). 28.002 PORT SF B[13][I][D1000] 28.003 5520 2.50V S 2W 28.004 PORT [D2000]QM[13][I][I][I] 28.005 MEME 28.006 MEMC V 0.0015U 29.001 5520 * S 29.002 HEAD {CAPACITANCE TESTS} # Select Capactance (Soft Key 5). 29.003 PORT SF C[13][I] 29.004 HEAD CAPACITANCE TESTS: {1uF Range} 29.005 5520 0.950uF S 2W 29.006 PORT [D2000]QM[13][I][I][I] 29.007 MEME 29.008 MEMC 1 uF 0.020U 30.001 5520 * S 30.002 DISP Disconnect the 5520A from the UUT. 30.003 HEAD CAPACITANCE TESTS: {10000pF Range} 30.004 ASK- U 30.005 MATH MEM1 = 0.0 30.006 PORT [D2000]QM[13][I][I][I] 30.007 MEME 30.008 MEMC 10000 pF 130.0U #! T.U.R. not calculated because ASK 'U' is not set. 31.001 ASK+ U 31.002 HEAD {RESISTANCE TESTS} 31.003 DISP Connect the 5520A to the UUT as follows: 31.003 DISP [32] 5520A AUX HI to UUT Volt Ohm Diode Input 31.003 DISP [32] 5520A AUX LO to UUT COM Input 31.003 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode Input 31.003 DISP [32] 5520A NORMAL LO to UUT COM Input 31.004 DISP Rotate the UUT function switch to Ohms. 31.004 DISP 31.004 DISP Select Advance when the numerical display appears. 31.005 HEAD RESISTANCE TESTS: {300 Ohm Range} 31.006 5520 0.00Z S CW # Press the RANGE button. 31.007 PORT SF 0[13][I][D1000] # Select Meter Manual (Soft Key 1). 31.007 PORT SF 8[13][I][D1000] 31.008 PORT [D2000]QM[13][I][I][I] 31.009 MEME 31.010 MEMC 300 Z +0.10U 32.001 5520 290.00Z S CW 32.002 PORT [D2000]QM[13][I][I][I] 32.003 MEME 32.004 MEMC 300 Z 0.30U 33.001 HEAD RESISTANCE TESTS: {3 kOhm Range} # Press Range up (Soft Key 2) to select the 3 kOhm range. 33.002 PORT SF 9[13][I] 33.003 5520 2.9000kZ S CW 33.004 PORT [D2000]QM[13][I][I][I] 33.005 MEME 33.006 MEMC 3 kZ 0.0022U 34.001 HEAD RESISTANCE TESTS: {30 kOhm Range} # Press Range up (Soft Key 2) to select the 30 kOhm range. 34.002 PORT SF 9[13][I] 34.003 5520 29.000kZ S CW 34.004 PORT [D2000]QM[13][I][I][I] 34.005 MEME 34.006 MEMC 30 kZ 0.022U 35.001 5520 * S 35.002 HEAD RESISTANCE TESTS: {300 kOhm Range} # Press Range up (Soft Key 2) to select the 300 kOhm range. 35.003 PORT SF 9[13][I] 35.004 DISP Connect the 5520A to the UUT as follows: 35.004 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode Input 35.004 DISP [32] 5520A NORMAL LO to UUT COM Input 35.005 5520 290.00kZ S 2W 35.006 PORT [D2000]QM[13][I][I][I] 35.007 MEME 35.008 MEMC 300 kZ 0.22U 36.001 HEAD RESISTANCE TESTS: {3 MOhm Range} # Press Range up (Soft Key 2) to select the 3 MOhm range. 36.002 PORT SF 9[13][I] 36.003 5520 2.9000MZ S 2W 36.004 PORT [D2000]QM[13][I][I][I] 36.005 MEME 36.006 MEMC 3 MZ 0.0046U 37.001 HEAD RESISTANCE TESTS: {30 MOhm Range} # Press Range up (Soft Key 2) to select the 30 MOhm range. 37.002 PORT SF 9[13][I] 37.003 5520 10.000MZ S 2W 37.004 PORT [D2000]QM[13][I][I][I] 37.005 MEME 37.006 MEMC 30 MZ 0.023U 38.001 HEAD {CONDUCTANCE TESTS} # Press the RANGE button. 38.002 PORT SF 0[13][I][D1000] # Select nS (Soft Key 3). 38.003 PORT SF A[13][I][D1000] 38.004 HEAD CONDUCTANCE TESTS: {300nS Range} 38.005 5520 285.71nY S 2W 38.006 PORT [D2000]QM[13][I][I][I][I$] 38.007 MEME 38.008 MEMC nY 1.63U 39.001 HEAD CONDUCTANCE TESTS: {3000nS Range} 39.002 5520 2857.1nY S 2W 39.003 PORT [D2000]QM[13][I][I][I][I$] 39.004 MEME 39.005 MEMC nY 16.3U 40.001 5520 * S 40.002 DISP Disconnect the 5520A from the UUT. 40.003 HEAD {LOGIC TEST} 40.004 DISP Connect the PM 5139 to the UUT EXT TRIG and COM 40.004 DISP Inputs using a BNC cable and a BNC to dual banana 40.004 DISP adapter. 40.005 DISP Rotate the UUT function switch to LOGIC. 40.005 DISP 40.005 DISP Select Advance when the numerical display appears. # Select TTL (Soft Key 1). 40.006 PORT SF 8[13][I][D1000] 40.007 DISP Setup the PM 5139 for a 2.0MHz, 1.5Vpp square, 40.007 DISP with 0.7Vdc offset. 40.008 ACC 2.0000MH TOL 40.009 PORT [D2000]QM[13][I][I][I][I$] 40.010 MEME 40.011 MEMC MH 0.0011U 1.5Vpp 41.001 HEAD {EXTERNAL TRIGGER} 41.002 DISP Rotate the UUT function switch to mVDC. 41.002 DISP 41.002 DISP Select Advance when the numerical display appears. # Press DISPLAY MODE. 41.003 PORT SF 3[13][I][D1000] # Press View (Soft Key 3). 41.004 PORT SF A[13][I][D1000] # Press Trigger (Soft Key 2). 41.004 PORT SF 9[13][I][D1000] # Select External (Soft Key 4). 41.004 PORT SF B[13][I][D1000] # Press Exit (Soft Key 5). 41.004 PORT SF C[13][I][D1000] # Press Single Shot (Soft Key 3). 41.004 PORT SF A[13][I][D1000] # Press Arm (Soft Key 1). 41.004 PORT SF 8[13][I][D1000] 41.005 DISP Setup the PM 5139 for 2.0Vpp, 100kHz Pulse, 50% duty. 41.006 EVAL Did the UUT trigger? # Press Exit (Soft Key 5). 42.001 PORT SF C[13][I][D1000] 42.002 HEAD {GLITCH CAPTURE} 42.003 DISP Move the PM 5139 connection to the UUT Volt Ohm Diode 42.003 DISP and COM Inputs. # Press DISPLAY MODE. 42.004 PORT SF 3[13][I][D1000] # Press View (Soft Key 3). 42.005 PORT SF A[13][I][D1000] # Press Glitch Capture (Soft Key 4). 42.005 PORT SF B[13][I][D1000] # Press Arm (Soft Key 1). 42.005 PORT SF 8[13][I][D1000] 42.006 DISP Setup the PM 5139 for 0.07Vpp, 10kHz Pulse, 50% duty. 42.007 EVAL Did the UUT trigger? 43.001 HEAD {DUTY CYCLE TEST} 43.002 DISP Rotate the UUT function switch to VAC. 43.002 DISP 43.002 DISP Select Advance when the numerical display appears. # Press the HZ/% button. 43.003 PORT SF 5[13][I][D1000] # Select Duty Cycle (Soft Key 2). 43.003 PORT SF 9[13][I][D1000] 43.004 DISP Setup the PM 5139 for 0.2Vpp, 10kHz square, 50% duty. 43.005 ACC 50.0pct 0.1% 43.006 PORT [D2000]QM[13][I][I][I][I$] 43.007 MEME 43.008 MEMC pct 0.4U 44.001 HEAD {FREQUENCY TESTS} # De-Select Duty Cycle (Soft Key 2). 44.002 PORT SF 9[13][I][D1000] 44.003 HEAD FREQUENCY TESTS: {Autorange} 44.004 DISP Setup the PM 5139 for 800kHz, 0.21Vpp square. 44.005 ACC 800.00kH TOL 44.006 PORT [D7000]QM[13][I][I][I][I$] 44.007 MEME 44.008 MEMC A kH 0.41U 0.21Vpp 45.001 DISP Disconnect the PM 5139 from the UUT. 45.002 DISP Connect the 5520A to the UUT as follows: 45.002 DISP [32] 5520A NORMAL HI to UUT Volt Ohm Diode Input 45.002 DISP [32] 5520A NORMAL LO to UUT COM Input 45.003 5520 3.00H 0.060V SI S 2W 45.004 PORT [D7000]QM[13][I][I][I][I$] 45.005 MEME 45.006 MEMC A H 0.02U 0.060V 46.001 5520 500.00kH 0.125Vpp SI S 2W 46.002 PORT [D7000]QM[13][I][I][I][I$] 46.003 MEME 46.004 MEMC A kH 0.26U 0.125Vpp 47.001 5520 1.5000MH 2.1V SI S 2W 47.002 PORT [D7000]QM[13][I][I][I][I$] 47.003 MEME 47.004 MEMC A MH 0.0009U 2.1V 48.001 5520 * S 48.002 DISP Connect the 5520A to the UUT as follows: 48.002 DISP [32] 5520A AUX HI to UUT mA/uA Input 48.002 DISP [32] 5520A AUX LO to UUT COM Input 48.003 DISP Rotate the UUT function switch to mA/uA. 48.003 DISP 48.003 DISP Select Advance when the numerical display appears. # Select AC (Soft Key 5). 48.004 PORT SF C[13][I][D1000] 48.005 HEAD FREQUENCY TESTS: {10kHz Range} # Press the HZ/% button. 48.006 PORT SF 5[13][I][D1000] 48.007 5520 10.000kH 3mA SI S 2W 48.008 PORT [D2000]QM[13][I][I][I][I$] 48.009 MEME 48.010 MEMC 10 kH 0.006U 3mA # Press Exit (Soft Key 5). 49.001 PORT SF C[13][I][D1000] 49.002 HEAD {DC CURRENT TESTS} # Select DC (Soft Key 5). 49.003 PORT SF C[13][I][D1000] 49.004 HEAD DC CURRENT TESTS: {30mA Range} 49.005 5520 21.000mA S 2W 49.006 PORT [D2000]QM[13][I][I][I][I$] 49.007 MEME 49.008 MEMC 30 mA 0.026U 50.001 HEAD {RMS AC CURRENT TESTS} # Select AC (Soft Key 5). 50.002 PORT SF C[13][I][D1000] 50.003 HEAD RMS AC CURRENT TESTS: {30mA Range} 50.004 5520 21.000mA 10kH SI S 2W 50.005 PORT [D2000]QM[13][I][I][I][I$] 50.006 MEME 50.007 MEMC 30 mA 0.440U 10kH 51.001 5520 * S 51.002 HEAD {DC CURRENT TESTS} 51.003 DISP Connect the 5520A to the UUT as follows: 51.003 DISP [32] 5520A AUX HI to UUT A Input 51.003 DISP [32] 5520A AUX LO to UUT COM Input 51.004 DISP Rotate the UUT function switch to A. 51.004 DISP 51.004 DISP Select Advance when the numerical display appears. 51.005 HEAD DC CURRENT TESTS: {3A Range} 51.006 5520 2.9000A S 2W 51.007 PORT [D2000]QM[13][I][I][I][I$] 51.008 MEME 51.009 MEMC 3 A 0.0073U 52.001 HEAD {RMS AC CURRENT TESTS} # Select AC (Soft Key 5). 52.002 PORT SF C[13][I][D1000] 52.003 HEAD RMS AC CURRENT TESTS: {3A Range} 52.004 5520 2.9000A 3kH SI S 2W 52.005 PORT [D2000]QM[13][I][I][I][I$] 52.006 MEME 52.007 MEMC 3 A 0.0228U 3kH 53.001 HEAD RMS AC CURRENT TESTS: {10A Range} # Press the RANGE button. 53.002 PORT SF 0[13][I][D1000] # Press Range up (Soft Key 2) to select the 10A range. 53.002 PORT SF 9[13][I][D1000] 53.003 5520 1.000A 3kH SI S 2W 53.004 PORT [D2000]QM[13][I][I][I][I$] 53.005 MEME 53.006 MEMC 10 A 0.018U 3kH 54.001 HEAD {AVG AC CURRENT TESTS} # Press the RANGE button. 54.002 PORT SF 0[13][I][D1000] # Select Average (Soft Key 3). 54.003 PORT SF A[13][I][D1000] 54.004 HEAD AVG AC CURRENT TESTS: {3A Range} 54.005 5520 2.900A 3kH SI S 2W 54.006 PORT [D3000]QM[13][I][I][I][I$] 54.007 MEME 54.008 MEMC 3 A 0.032U 3kH 55.001 5520 * S 55.002 DISP Disconnect the 5520A from the UUT. 55.003 HEAD {COMPONENT TEST} 55.004 DISP Connect the Fluke 85 to the UUT as follows: 55.004 DISP [32] 85 VOLT OHM DIODE to UUT Volt Ohm Diode 55.004 DISP [32] 85 COM to UUT COM 55.005 DISP Rotate the Fluke 85 function switch to VAC. 55.005 DISP Rotate the UUT function switch to COMPONENT TEST. 55.006 ACC 2.85V 1% 0.002U 55.007 MEMI Enter Fluke 85 reading in volts: 55.008 MEMC 2.85V 0.65U 200H 56.001 DISP Connect the Fluke 85 to the UUT as follows: 56.001 DISP [32] 85 mA uA to UUT Volt Ohm Diode 56.001 DISP [32] 85 COM to UUT COM 56.002 DISP Rotate the Fluke 85 function switch to uAAC. 56.003 ACC 210uA 0.6% 0.2U 56.004 MEMI Enter Fluke 85 reading in microamps: 56.005 MEMC 210uA 46U 200H 57.001 END